CR10XTCR Thermocouple Reference for CR10X Wiring Panel

Überblick

Please note, both the CR10X and CR10XTCR have been retired. The CR10XTCR provides a temperature reference for thermocouple measurements.

The CR10XTCR is compatible with the black CR10X wiring panel only (shipped after 21 May 97). Refer to Compatibility for more information.

Lesen Sie mehr

Bilder

Technische Beschreibung

The CR10XTCR provides a temperature reference for thermocouple measurements. Its leads connect directly to channels 1H, E3, and AG on the CR10X Wiring Panel. The thermistor and resistors are attached to a small printed circuit board (pn 8289).

The aluminum thermocouple cover (pn 10630) reduces thermal gradients across the analog inputs.

Spezifikationen

Temperature Range -35° to +50°C
Typical Polynomial Linearization Accuracy
  • < ±0.5°C over full temperature range
  • < ±0.1°C over -24° to +48°C range
Typical Interchangeability Error < ±0.2°C over 0° to 50°C range (increasing to ± 0.5°C at -40°C)

Kompatibel mit

The CR10XTCR is compatible with the black CR10X wiring panel only (shipped after 21 May 97).

The CR10XTCR is NOT compatible with the black CR10 wiring panel or the silver wiring panel. These wiring panels use either the CR10TCR or 10TCRT (see Related Products).