How to Easily Replace and Update a CS223 Sensor Head

von Ajay Singh | Aktualisiert: 01/15/2024 | Kommentare: 0

Blog Themen


Suchen Sie im Blog


Blog abonieren

Erhalten Sie eine Email wenn ein neuer Artikel veröffentlicht wird. Wählen Sie die Themen die Sie interessieren.


Geben Sie Ihre Emailadresse ein:



Artikel vorschlagen

Gibt es ein Thema zu dem Sie mehr erfahren möchten?

Leave this field empty

CS223 and CS241

Campbell Scientific’s CS223 Pt-1000 Class A, Back-of-Module Temperature Sensor (discontinued effective 31 October 2023) has been replaced by the CS241 Pt-1000 Class A, Back-of-Module Temperature Sensor. When needed, you can replace the CS223 sensor head with a CS241 head without having to replace the entire cable. This blog article explains the steps to take so you can use the existing cable already installed in the field.

Configurations for the CS223

The CS223 was available in two configurations. Check to see which one you have, as the instructions for replacement are somewhat different.

CS223-L-PW

This configuration used prewired connectors, which was most preferred by our customers as it minimized the need for wiring in the field. The configuration consisted of three parts:

  1. A sensor head with a small cable with a Q-Disconnect connector (pn 19520)
  2. A home run cable connecting the sensor with the enclosure containing the data logger. The cable had pn 26146 D-Disconnect mating connector to the part in (1) above and another connector (pn 22018) that mated with a bulkhead connector mounted on the enclosure.
  3. The bulkhead connector (pn 22019) mounted on the enclosure that had a high-precision 1,000 ohm (pn 19160) completion resistor soldered on it. All necessary connections between this connector and the data logger are done at the factory.



CS223-L-PW configuration

CS223-L-TIM

This version used the 3WHB10K Terminal Input Module (TIM) for the completion resistor and did not use a bulkhead panel-mount connector. The customers had to complete the field wiring themselves. The parts in 1 and 2 above were the same. The parts in 3 were not used. A wiring diagram is shown below.



CS223-L-TIM configuration

Converting the CS223 Sensor Head to a CS241 Sensor Head

For the conversion, you will need pn 32932 and a circular, plastic, M12-type socket, eight-pin connector sold as a kit (pn 39570), also referred to as the CS241 field install kit for CS240 replacement.

You will need this if you want to use the existing cabling but want to replace the CS223 with a CS241.

Steps for CS223-L-PW

  1. Cut the Q-Disconnect connector from the end facing the sensor. This is shown in the red circle in the diagrams above. Please cut the connector as close to the end as possible. You can also watch our M12 Connector Field Install Kit video
  2. Remove the outer insulation jacket carefully to about 1”.
  3. Strip the individual wires about 1/4”.
  4. Connect the wires to the connector pins as follows:

    Wire Color Pin on the Connector Data Logger
    White 1
    Blue 3 SE-L
    Gray 4 SE-H
    Black 5 Excite (Vx1)

  5. Cut off the connector on the other end of the cable and land the conductors directly to the data logger. (See the table above.)

Steps for CS223-L-TIM

  1. Cut the Q-Disconnect connector from the end facing the sensor. This is shown in the red circle in the diagrams above. Please cut the connector as close to the end as possible. You can also watch our M12 Connector Field Install Kit video.
  2. Remove the outer insulation jacket carefully to about 1”.
  3. Strip the individual wires about 1/4”.
  4. Connect the wires to the connector pins as follows:

    Wire Color Pin on the Connector Data Logger
    White 1
    Blue 3 SE-L
    Gray 4 SE-H
    Black 5 Excite (Vx1)
  5. Remove the TIM module and connect the wires to the respective data logger channels directly.
  6. Install a jumper between the Excite channel and SE-H channel used. You don’t need the TIM module, as the CS241 comes with a completion resistor installed inside the M12 connector.

Please use the following lines of CRBasic code for a CR1000X Measurement and Control Datalogger to read the sensor:


Public RRTD
Public BomT
DataTable (Test,True,1000)
	DataInterval (0,1,Sec,10)
		Sample (1,BomT,FP2)
EndTable 

BeginProg
	Scan (1,Sec,0,0)
		BrHalf3W (RRTD,1,mV1000,1,Vx1,1,500,True ,0,60,1000,0)
		PRTCalc (BomT,1,RRTD/1000,1,1.0,0)
	CallTable Test

NextScan
  
EndProg

Final Words

In case you have some other variation, you can still use the methods outlined here and a good digital handheld multimeter to troubleshoot, just in case. You can also reach out to us by contacting our support staff, and we will be happy to help you.